Ageing (Durability) Test Bench
Product Description
A PLC-controlled ageing test bench that performs long-term durability (endurance) and wear tests on water meters in line with EN ISO 4064 and OIML R49. Meters with Q3 ≤ 16 m³/h undergo 100,000 cycles of intermittent flow and 100 hours of continuous flow at Q4; large meters with Q3 > 16 m³/h undergo 800 hours of continuous flow at Q3 and 200 hours at Q4. Durability is assessed by comparing the indication error before and after the test.
Technical Specifications
- Standards
- EN ISO 4064, OIML R49-2:2024 (clause 7.11)
- Meter sizes
- DN15–DN40 and DN50–DN∞
- Intermittent flow
- 100,000 cycles · 15 s flow / 15 s stop
- Continuous flow (Q3 ≤ 16 m³/h)
- 100 h at Q4
- Continuous flow (Q3 > 16 m³/h)
- 800 h at Q3 · 200 h at Q4
- Test water (T30 / T50)
- 20 ± 5 °C
- Control
- PLC + touch panel
- Reporting
- PDF / Excel
Water Meter Ageing and Durability Test Bench
DN15–DN40 | Continuous and intermittent flow test system
EN ISO 4064 and OIML R49 Compliant
Endurance and durability tests of water meters performed in line with international standards.
DN15–DN40 Test Capacity
Modular test line suitable for mechanical, ultrasonic and electronic water meters.
100,000-Cycle Intermittent Flow
Automatic durability test based on 15 s flow and 15 s stop for meters with Q3 ≤ 16 m³/h.
100-Hour Continuous Flow Test
100-hour durability test at the Q4 overload flow rate for meters with Q3 ≤ 16 m³/h.
Automatic Flow Control
Stable flow and controlled flow transitions with frequency-controlled pumps and automatic valves.
Automatic PLC Test Management
Automatic management of run, stop, cycle count, flow and test durations.
Multi-Meter Test Line
Several water meters are aged at the same time in series or parallel connection.
Precision Reference Flowmeter
The flow rate is monitored throughout the test and kept within the standard tolerances.
Temperature and Pressure Control
Continuous monitoring of test water temperature and inlet and outlet pressures.
Water Hammer and Cavitation Protection
Controlled valve movements and suitable line pressure prevent additional wear on the meters.
Touch Control and Monitoring
Display of remaining test time, cycle count, instantaneous flow, pressure and temperature.
Automatic Logging and Reporting
Test duration, flow, cycle records, alarm history and results are reported as PDF / Excel.
Large Water Meter Ageing and Durability Test Bench
DN50 and above | High-flow continuous flow test system
EN ISO 4064 and OIML R49 Compliant
Standard-compliant design for long-term durability and wear tests of large water meters.
DN50 and Larger Test Capacity
Modular test setup for flanged Woltmann, ultrasonic and electromagnetic meters.
800-Hour Continuous Flow Test at Q3
800-hour long-term durability test at permanent flow for meters with Q3 > 16 m³/h.
200-Hour Overload Test at Q4
Meters with Q3 > 16 m³/h are run for 200 hours at overload flow to assess their durability.
High-Capacity Pump System
Frequency-controlled pump infrastructure matching the Q3 and Q4 flow requirements of large meters.
Automatic PLC Test Control
Automatic management of flow setting, test duration, pump control and test safety.
Flanged, Modular Connection
Interchangeable connection adapters for different meter sizes and pressure ratings.
Precision Reference Flowmeters
Verification of stable flow and continuous flow monitoring over a wide flow range.
Continuous Pressure and Temperature Monitoring
Real-time logging of meter inlet/outlet pressures and test water temperature.
Cavitation and Water Hammer Protection
Controlled flow changes, pressure protection and safe valve control.
24/7 Uninterrupted Test Monitoring
Automatic alarms, data logging and safe shutdown for long-duration tests.
Before / After Test Comparison
Changes in indication error are analysed and durability results reported as PDF / Excel.
Endurance Test Durations According to the Standards
OIML R49-2:2024 clause 7.11 – Table 1
| Test condition | Q3 ≤ 16 m³/h | Q3 > 16 m³/h |
|---|---|---|
| Intermittent flow | 100,000 cycles | Not applied to normal meters |
| Flow rate | Q3 | Q3 and Q4 |
| Run / stop | 15 s / 15 s | — |
| Continuous test at Q3 | — | 800 h |
| Continuous test at Q4 | 100 h | 200 h |
| T30 / T50 test water | 20 ± 5 °C | 20 ± 5 °C |
For other temperature classes, 50 ± 5 °C applies to the Q3 test and 0.9 × MAT ± 5 °C to the Q4 test. Additional durability tests apply to combination water meters.